Dust and environment monitors · Monitoring instrument
Indoor air quality monitors

Designed for gas and particle pollutant measurements to better understand atmospheric chemistry and environmental public health, these internet-enabled air quality monitors provide a near-instantaneous look at changing pollutant concentrations, with peripheral sensors for measuring environmental and meteorological parameters.
At a glance
- Particle sizes
- 0.3–25 µm, 6 channels
- PM fractions
- PM1, PM2.5, PM5, PM10
- Other parameters
- CO₂, temperature, RH; TVOC (5302-AQM)
- Flow rate
- 0.1 CFM (2.83 L/min)
- Connectivity
- Ethernet, USB; optional 802.11 b/g, RS485, RS232
- Memory
- 45,000 records (rotating buffer)
- Power
- 110–240 VAC, 50/60 Hz
- Operating conditions
- 5–40 °C · 20–95% non-condensing
Taken from the technical text ABCDust publishes, section by section.
Downloadable documents
Models 5301-AQM and 5302-AQM (CO₂, temperature, RH and TVOC)
- The 5301-AQM and 5302-AQM Remote Air Quality Monitors measure and display particle mass concentrations for PM1, PM2.5, PM5 and PM10, as well as temperature, relative humidity and CO₂ (and TVOC in the 5302-AQM). This wall-mounted instrument is designed for fixed installations. Its advanced power management and sleep mode allow long intervals between samples over extended, unattended operation. The AQM Series can easily be integrated into a building automation and facility monitoring system via Ethernet, USB or (optional) wireless 802.11 b/g, RS485 or RS232.
- Both models report and display 6 user-selectable particle size channels, as well as carbon dioxide (CO₂), temperature and relative humidity. The instrument monitors PM1, PM2.5, PM5 and PM10 (and any other PM size value the user specifies) with an easy-to-use Channel Management control panel. The instrument also monitors flow, sensor status, date of last calibration, serial number, laser on-time and date of manufacture for easy maintenance and warranty management. Furthermore, the remote monitor operates using a long-life internal 0.1 CFM pump.
- You can also view data and generate environmental reports in real time on screen, via printer or USB key, through the output options, or export them to the Instrument Management Software during a monitoring system or network failure. The units can be controlled and monitored remotely via a web browser from any PC, tablet or smartphone, supporting up to 20 concurrent user sessions.
- Moreover, the 5301-AQM and 5302-AQM particle counters store 45,000 time-stamped particle count and environmental data records for redundant protection in the event of a monitoring system or network failure. The AQM Series mass concentration mode approximates density in µg/m³ and PM (particulate matter), allowing for density and refractive index corrections to ensure accuracy. The units meet ISO 21501-4 and JIS B9921. Finally, these instruments ensure compliance with an on-board pulse height analyzer.
Features and benefits
- Measures mass concentration in µg/m³ and indicates PM (particulate matter)
- Measures up to 6 channels of simultaneous data
- Measures CO₂, temperature and relative humidity with user alarm set points
- Measures TVOC in parts per billion (Model 5302-AQM)
- User-adjustable single- and two-point sensor calibration offset
- User-selectable channel sizes
- Long-life laser diode technology
- Displays and externally prints environmental reports
- Internal audible alarm with user-selectable thresholds for all environmental parameters
- User-definable reports
- Easy to clean and wipe down, with minimal particle traps
Specifications
- Size range: 0.3 to 25 µm
- Size channels: factory calibrated at 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
- Flow rate: 0.1 CFM (2.83 LPM)
- Concentration limit: 4,000,000 particles/ft³ at 5% coincidence loss
- Optional communication modes: wireless 802.11 b/g, RS485 or RS232
- Environmental sensor: optional temperature and relative humidity probe, 32 °F to 122 °F (0 °C to 50 °C) ±1 °F (0.5 °C), 15–90% ±2% relative humidity
- Alarm: alarms on counts for all particle sizes, sensor failure, environmental sensors and flow
- Standards: ISO 21501-4 and JIS B9921
- Calibration: recommended minimum once per year
- External surface: stainless steel
- Dimensions (L × W × H): 5.2 in × 2.3 in × 8.9 in (13.3 cm × 5.8 cm × 22.6 cm), including barb fittings
- Weight: 2.6 lb (1.18 kg)
Accessories
Quick start guide, operating manual on USB flash drive, isokinetic probe, power supply and cable
Optional accessories
Printed manual, temperature/relative humidity sensor, isokinetic probes, purge filter, data cable, mounting bracket and sample tubing
Buffer memory
- 45,000 sample records (rotating buffer) including particle count data and environmental data
- Sample time: 1 second to 99 hours
- Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
- Operating conditions: 41 °F to 104 °F (5 °C to 40 °C) / 20% to 95% non-condensing
- Storage conditions: 32 °F to 122 °F (0 °C to 50 °C) / up to 98% non-condensing
- Warranty: 2 years. Extended warranties available.